Adapter Output: DC 15V, 2.0A
Adapter Output: DC 15V, 2.0A
Compatible tube:Transfusion set, infusion set or feeding set with diameter 3.5-5mm
Adapter Output: DC 15V, 2.0A
Protection class:I
Heating board Degree of protection:BF
Main engine Protection against ingress of water:IPX2
Control temperature:35°C - 42°C adjustable in 0.1°C steps
Temperature accuracy:±0.5°C (heating plate), ±1°C (warming tube)
Operational mode:Continuous operationDimensions :140x75x35mm Weight:210g
service life:5 years
Tube warmer Length: 800mm
Permissible environmental conditions for transport and storage:Temperature:-20°C – 55°C
Relative humidity:10– 90%
Atmosphere range:70 – 106 Kpa
Permissible environmental conditions for the ambient operating:
Temperature: +5 – 40°C.
Values higher or lower than the ranges specified above may cause damage to the device or its accessories.
Relative Humidity: ≤80%
Atmosphere range: 70 – 106 Kpa
Temperature
Use the warming slot output reference table :
Infusion Rate(ml/h) | 25 | 100 | 180 | 300 | 600 | 1000 | |||||||
Slot Length | Part | All | Part | All | Part | All | Part | All | Part | All | Part | All | |
Temperature Set(℃) | 35 | 28.2 | 29.0 | 32.0 | 33.0 | 32.4 | 33.2 | 30.5 | 32.0 | 28.2 | 29.4 | 26.0 | 27.0 |
| 37 | 29.2 | 30.8 | 33.4 | 34.8 | 33.4 | 35.0 | 31.6 | 33.6 | 28.8 | 30.4 | 26.5 | 27.5 |
| 42 | 30.4 | 33.2 | 36.6 | 39.0 | 37.8 | 39.5 | 36.0 | 37.5 | 30.7 | 34.4 | 28.1 | 30.0 |
Note: Room temperature is 20 degrees, the temperature of the input liquid is 20 degrees, the output port of the warmer is at 125px from the outlet of the Warmer.
Use warmer tube output reference table:
Infusion Rate(ml/h) | 25 | 100 | 180 | 300 | 600 | 1000 | |
Temperature Set(℃) | 35 | 29.2 | 33.8 | 34.5 | 34.8 | 32.2 | 30.0 |
| 37 | 31.2 | 35.6 | 36.5 | 36.5 | 33.8 | 30.5 |
| 42 | 33.6 | 39.8 | 41.0 | 41.0 | 36.2 | 32.5 |
Note: Room temperature is 20 degrees, the temperature of the input liquid is 20 degrees, the output port of the warmer is at 125px from the outlet of the Warmer Tube.
The heating effect is on the basis of the actual use (infusion environment temperature, infusion rate, soft and hard, thickness, and diameter of the infusion tube wall will affect the heating effect)