When using nitrous oxide, oxygen concentration>25%;
When using nitrous oxide, oxygen concentration>25%;
LTSA05 Anesthesia Machine(10.4 Inches screen)
Main body:High-strength engineering plastic rack, light,
beautiful and corrosion resistant
Scope of application :adult, child
Gas source : O2: 0.27~0.55MPa N2O: 0.27~0.55MPa
Flow meter:O2: 0.1~1.0L/min 1.1~10L/min
N20: 0.1~1.0L/min 1.1~10L/min
O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
Flow rate of rapid oxygen supply:25~75L/min
Low oxygen pressure alarm :There will be sound alarm when the oxygen pressure < 0.2MPa
Vaporizer :
it has the function of automatic compensation based on pressure, temperature, and flow rate.
The regulation range of evaporator concentration is 0~5 vol%. Among Halothane,Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.
Respiratory circuit:working mode: all-close, semi-close, semi-open
releasing pressure: 0.5~7 kPa
Respiratory Bellows:bellows for adults, bellows for children
tidal volume range:0~1,500 ml
Ventilator
Display mode:High-definition TFT color LCD screen display
Ventilation mode :IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH, (IMV, VCV, PCV)
Ventilation parameters:
Tidal volume 50~1500 ml (20~1500 ml)
Rate 1~99 bpm
SIMV rate 2~20 bpm
I:E 2:1~1:8 (4:1~1:8)
Inspiratory trigger pressure -1.0~2.0Kpa
PEEP 0~2.0Kpa
Pressure Range 1.0~6.0Kpa
SIGH 1.5 times the inspiratory time (70-150 adjustable)
Inspiratory platform 0~50%
Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform (autonomous inspiratory rate, lung compliance, airway resistance,flow
-volume oop,pressure-volume loop)
Monitoring of oxygen concentration 21%~100%
Security Alarm System
Oxygen Concentration alarm:
upper limit setting range 21%~100%
low limit setting range 10%~80%
Airway pressure alarm:
upper limit setting range 0.3~6.0 kPa
low limit setting range 0.2~5.0 kPa
Per-minute ventilation volume alarm
upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min
Sustained high-pressure alarm : it will give alarm when stress have consistently been higher than 2.5 kPa
Suffocation alarm:it will give sound and light alarm if there is no tidal volume input for 15 seconds
Power alarm
Power :AC 220V±10% 50Hz±1Hz(Optional UPS, Storage battery)
(Mechanical Arms) (End-expiratory CO2 monitoring)
*Items within brackets are optional.
Q:What's the minimum order quantity (MOQ)?
A:For most of our medical products, even order for only one unit is warmly welcomed.
Q: Can you do OEM/ private label?
A: Of course we can do OEM/private label for you.
Q: What's your delivery time?
A: Generally it is 10 days if the goods are in stock.
or it is 15-20 days if the goods are not in stock, it is according to quantity.
Q: How to ship the order?
A: Please inform us your instruction, by sea, by air or by express, any way is ok for us. We have very professional forwarder to provide the best shipping cost, service and guarantee.
Q: What is your terms of payment ?
A: We accept T/T ,LC,Western Union ,Paypal and more.Please suggest your preferred payment method.