LTCM03V Vet Microplate Reader
  • LTCM03V Vet Microplate Reader

  • No: LTCM03V
  • Type:
  • Dimension:
  • Characteristics:

    Filter choosing-mode: automatically


Details

LTCM03V Vet Microplate Reader

Packing Details: 520*410*255mm,  0.06CBM;  NW: 10KG,   GW:13KG

Elisa Reader, 8 channels optical system, 5 secs/96 wells(single wavelength)some are 6 secs, 10 secs /96 wells ,.7 secs/96wells(double wavelength )some products are 20 s , 4 filters,CE marke

 

Specifications

 Measurement System  :  8-Channel optical system
 Measurement Range   :  0.001~4.000Abs
 Photometric Accuracy :  ±0.01Abs  (0.001~1.500Abs)
Repeatability :  ±0.5% (0.001~1.500Abs)
Stability :  0.005Abs 
Linearity :  ±1% (0.001~1.500Abs)
Filters :  Standard:4 filters (405,450,492,630nm) ,other wave length can be optional.
Totally 7 filters can be installed.

Full One year Warranty

  Characters

The Microplate Reader is a semiautomated benchtop instrument focused on clinical immunity-test in hospitals or epidemic preventing stations for Vet. It has good characters such as high sensitivity, high accuracy, high speed of detection and the easiness of operation etc. Display is a liquid crystal one with alphanumeric and graphic options. Parameters can be checked on the display without using difficult command sequences.

 Main Functions

The instrument is compatible with registered diagnostic EIA kit, primarily used in hospitals or some detections related to clinical immunology.

  Function and Technical Specifications

Fundamental parameters

1) Range of working wavelength is from 400nm to 800nmRange of detection is from 0.000A to 4.000A

2) Resolving power0.001 A

3) Filters when leaving factory405nm450nm492nm630 nm, with the maximum array of 10 filters between 400nm-800nm simultaneously.

4) Reading speedsingle wave :5sdual waves :10s

5) Dimension: 420×325×210 mm(l×h×w)

6) Weight: 9 kg

7)Monitor240*128mm  LCD

8Printerexternal printer.

          Technical Specifications

(1) Characters of the filters

Table 1: Characters of the Filters

Accuracy of  wavelength  (nm)

Half-widthnm

±3.0

≤12

(2 )  The errors of absorbance accuracy are as follows:

Table 2: Absorbance Accuracy

Absorbance (A)

Accuracy Error

0.000-2.000

±0.01

2.000-3.000

±0.05

3.000-4.000

±0.15

(4) The Linearity Errors are as follows:

Table 3: Linearity of the Instrument

Absorbance (A)

Error (%)

0.000-2.000

±1.0%

2.000-3.000

±2.0%

3.000-4.000

±5.0%

(5) Reproducibility Error: ≤0.5%;

(6) Stability Error: ≤±0.005A/10min

Main function

1) Provided with power on self-checking. Error data will display on LCD monitor if something wrong with the photoelectric system.

2Filter choosing-mode: automatically

3) Detection mode: single wave/dual wave detection; single aperture /dual aperture detection

4Imput mode:key in film keyboard;

5Imput format: imput capital letter/lowercase letter/number at random;

6Item parametersProvided with function of set/save/ modify/ inquiry 126 calibration parameters and curves, to make detecting course more conveniently.

7Defection of calibration: with the function of calibration-defectionwhich can save/modify 126 parameters of calibration and curves.

Detection methodsabsorbance  /qualitative /quantitative  detection.

Qualitative detection: single critical value checkingreset gray area and weak positive aperture controlling;

Quantitative detection method: single-point, point-point, linear regression, semi-logarithm regression, logarithm regression,ratio regressionratio-semi-logarithm regression, exponential regression, quadratic-curves.

9Saving functioncan save the detection results on 105 pieces of plates

10Output mode: Absorbancedisplay/printout originally absorbable value

    Qualitative checkingdisplay/printout absorbable value, critical value, S/CO, qualitative results;

    Quantitative checkingdisplay/printout absorbable value, density, standard reference valuequantitative results.

11QC functioninquiry/printout the QC data and chart

12Communicationequipped with RS-232 interface.

13With self-monitoring functionreal-timeon each testing light-source signal channel.

14With dormancy function on light-source lamp.

 

 