LTCM03V Vet Microplate Reader
Packing Details: 520*410*255mm, 0.06CBM; NW: 10KG, GW:13KG
Elisa Reader, 8 channels optical system, 5 secs/96 wells(single wavelength)—some are 6 secs, 10 secs /96 wells ,.7 secs/96wells(double wavelength )—some products are 20 s , 4 filters,CE marke
Specifications |
Measurement System : 8-Channel optical system Full One year Warranty |
Ⅰ Characters
The Microplate Reader is a semiautomated benchtop instrument focused on clinical immunity-test in hospitals or epidemic preventing stations for Vet. It has good characters such as high sensitivity, high accuracy, high speed of detection and the easiness of operation etc. Display is a liquid crystal one with alphanumeric and graphic options. Parameters can be checked on the display without using difficult command sequences.
Ⅱ Main Functions
The instrument is compatible with registered diagnostic EIA kit, primarily used in hospitals or some detections related to clinical immunology.
Ⅲ Function and Technical Specifications
Fundamental parameters
1) Range of working wavelength is from 400nm to 800nm; Range of detection is from 0.000A to 4.000A;
2) Resolving power:0.001 A;
3) Filters when leaving factory:405nm,450nm,492nm,630 nm, with the maximum array of 10 filters between 400nm-800nm simultaneously.
4) Reading speed:single wave :5s,dual waves :10s;
5) Dimension: 420×325×210 mm(l×h×w);
6) Weight: 9 kg;
7)Monitor:240*128mm LCD;
8)Printer:external printer.
Ⅳ Technical Specifications
(1) Characters of the filters
Table 1: Characters of the Filters
Accuracy of wavelength (nm) | Half-width(nm) |
±3.0 | ≤12 |
(2 ) The errors of absorbance accuracy are as follows:
Table 2: Absorbance Accuracy
Absorbance (A) | Accuracy Error |
0.000-2.000 | ±0.01 |
>2.000-3.000 | ±0.05 |
>3.000-4.000 | ±0.15 |
(4) The Linearity Errors are as follows:
Table 3: Linearity of the Instrument
Absorbance (A) | Error (%) |
0.000-2.000 | ±1.0% |
>2.000-3.000 | ±2.0% |
>3.000-4.000 | ±5.0% |
(5) Reproducibility Error: ≤0.5%;
(6) Stability Error: ≤±0.005A/10min;
Ⅴ Main function
1) Provided with power on self-checking. Error data will display on LCD monitor if something wrong with the photoelectric system.
2)Filter choosing-mode: automatically
3) Detection mode: single wave/dual wave detection; single aperture /dual aperture detection
4)Imput mode:key in film keyboard;
5)Imput format: imput capital letter/lowercase letter/number at random;
6)Item parameters: Provided with function of set/save/ modify/ inquiry 126 calibration parameters and curves, to make detecting course more conveniently.
7)Defection of calibration: with the function of calibration-defection,which can save/modify 126 parameters of calibration and curves.
Detection methods:absorbance /qualitative /quantitative detection.
Qualitative detection: single critical value checking,reset gray area and weak positive aperture controlling;
Quantitative detection method: single-point, point-point, linear regression, semi-logarithm regression, logarithm regression,ratio regression,ratio-semi-logarithm regression, exponential regression, quadratic-curves.
9)Saving function:can save the detection results on 105 pieces of plates
10)Output mode: Absorbance:display/printout originally absorbable value;
Qualitative checking:display/printout absorbable value, critical value, S/CO, qualitative results;
Quantitative checking:display/printout absorbable value, density, standard reference value,quantitative results.
11) QC function:inquiry/printout the QC data and chart;
12) Communication:equipped with RS-232 interface.
13) With self-monitoring function(real-time)on each testing light-source signal channel.
14) With dormancy function on light-source lamp.