1、 Single-table/single-tube
2、 Rotary anode X-ray tube unit and tangential annular tube head
3、 Single-phase full wave rectification high-voltage generator
Prev:LTX23 X ray machine
Next:LTX21B x ray machine
USE
General radiography,Filter-ray-filter radiograph
Control, HT generator unit,Ht cable,X-ray tube unit,Radiographic table-X-ray source column ,Collimator
MAIN CHARACTERISTICS
1、 Single-table/single-tube
2、 Rotary anode X-ray tube unit and tangential annular tube head
3、 Single-phase full wave rectification high-voltage generator
4、 Power voltage(v) radiograph kilovolt (KV).step-less control electric mechanism
5、 Filament stabilizer for X-ray tube unit and space charge compel mentor
6、 Radiograph volume KV 、mA and S. grading and interlock protection
7、 Adopt digital circuit timer. Grading according to R10 priority coefficient,(for more accurately controlling time)
8、 HT primary, the zero controlled circuit of silicon-controlled rectifier of heavy power
9、 Radiographic table can move in length an breadth
10、The radiographic table. Column and vibrating grid are in a whole without ceiling and floor rails bottom
MIN TECHNICAL INDEXES
ITEM | CONTENTS | INDEX |
Demand of power supply | Capacity | Two-phase 30KVA |
Voltage、Frequency | 380V/220V ±10%50HZ ±0.5Hz | |
Intemal resistance | 0.752Ω(380V) 0.25Ω(220V) | |
power supply | Adjustable range | 380±10% at220V being adjustable continuously |
Radiograph | Voltage | 50-100KV 连续可调 50-100KV being adjustable continuously |
current | Small focus:50mA100mA,Big focus:50mA 100mA 200mA 300mA | |
time | 0.04~6.35s 23grades together grading a according to R10 coefficient | |
High-voltage generator | Capacity | 20KVA(instantaneous) |
Max DC output voltage | 100KV | |
Max DC output current | 200mA | |
X-Ray tube | Model | XD51.20-40/100 |
Focus | Small focus 1×1mm big focus2×2mm | |
Radiographic table | Length×width×height of bed surface | 2000mm×665mm×700mm |
Moving range of bed surface | 640mm in Length,200mmin breadth | |
Ray-filter of table | Travel in length ≥500mm | |
Grid density N=28 | ||
Convergence distance f。=100cm | ||
Grid ratio r=8 | ||
Column of X-ray tube unit | Moving Longitudinally along Radiographic table | 1000mm or 1800mm |
Distance to friend as moving up and down (along column) | 650~1800mm | |
Rotary around the center of cross arm | ±180° | |
Rotary around the shaft line of X-ray annular tubes | -10°~60°~ +120° | |
Max size of cassette | 356mm×432mm (14″×17″) |
r/>
Q:What's the minimum order quantity (MOQ)?
A:For most of our medical products, even order for only one unit is warmly welcomed.
Q: Can you do OEM/ private label?
A: Of course we can do OEM/private label for you.
Q: What's your delivery time?
A: Generally it is 10 days if the goods are in stock.
or it is 15-20 days if the goods are not in stock, it is according to quantity.
Q: How to ship the order?
A: Please inform us your instruction, by sea, by air or by express, any way is ok for us. We have very professional forwarder to provide the best shipping cost, service and guarantee.
Q: What is your terms of payment ?
A: We accept T/T ,LC,Western Union ,Paypal and more.Please suggest your preferred payment method.