Vet lab analyzer

Return

Vet lab analyzer

Home  > Products  > Veterinary equipment > Vet lab analyzer

Products

CUSTOMER CASE

LTCC04V Vet Semi auto chemistry analyzer
  • LTCC04V Vet Semi auto chemistry analyzer
  • LTCC04V Vet Semi auto chemistry analyzer

  • No: LTCC04V
  • Type: Vet lab analyzer
  • Dimension:
  • Characteristics:
  • Details
  • Pictures
  • Service
  • Q&A

  • Light Source: 6V 10W, halogen lamp of more than 2000 hours life time

    Spectrometer: interference filter.

    Wavelength: 340nm405nm505nm546nm578nm620nm, and one wavelength free.

    CML precision: ±2nm

    FWHM: ≤10nm

    Stray light≤0.5%(absorbance≥2.5)

    Absorbance range: -0.2143.000

    Absorbance linearity: ±1%

    Absorbance repeatability: ±0.001

    Absorbance resolution: 0.0001

    Precision: ≤1.0

    Carry over: ≤1%

    Temperature: 25℃30℃37℃,±0.3℃

    Flow cellpermanent quartz flow cell, minimal volume is 32μL, and disposable flow cell

               is selectable

    Sampling system: pressure auto releasing pumps ensure the accuracy of suction volume

    Operational system: VISTAWindows XPWindows2000

    Test Method: End point,Fixed time,Kinetic,Factor,Muli-standards,Bichromatic,Dual wavelengths

    Parameter Setting: Patients profile and calculated item can be set easily

    Absoption volume: 200μL3000μL

    Sample volume: 10μL50μL

    Reaction curve: real time display the reaction curve

    Memory: Able to store more than 200 parameters while 55 parameters of them are stablized ;

    Able to save more than 15000 test results and auto save the results if the power is off during working

    QC: auto draw the QC figure

    Reagent: Open reagent system

    Display: LCD with high brightness

    Input: press button

    Printer: Inner thermal printerable to be connected to external printer

    Report content: absorbanceconcentrationactivity, detailed patient information and

                  abnormal results marking

    Interface: RS-232 serial port, able to be connected to computer

    Working Condition: Temperature1030Relative Humidity20%80%

    Voltage: AC 85-264Vauto adapt according to the input voltage

    Power : ≤100W

    Quality System: Recognized by TUV and CMD

     

     

    Test item

    Liver

    ALTALPGGTTPAlbTBILDBIL

    Heart

    LDHCKCK-MBAST a-HBDH

    Lipids   

    CHOTGHDL_CLDL_C

    Lipoprotein

    Apo A1Apo BLpa

    kidney

    BUNCREUA

    Ion

    NaMgCaKPCl

    Other

    CO2-CPAMSLIPGLUIgAIgGIgMC3C4CRP

     

     

     

     

     