Ventilator
Display mode:High-definition 10.4” TFT color LCD screen display
Ventilation mode:IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH(IMV, VCV, PCV)
Ventilation parameters
Tidal volume:50~1500 ml (20~1500 ml)
Rate:1~99 bpm
SIMV rate:1~20 bpm
I:E :2:1~1:8 (4:1~1:8)
Inspiratory trigger pressure -1.0~2.0Kpa
PEEP :0~2.0Kpa
Pressure range:1.0~6.0Kpa
SIGH:1.5 times the inspiratory time(70-150 adjustable)
Inspiratory platform 0~50%
Parameters for ventilation monitoring:tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform (autonomous inspiratory rate, lung compliance, airway resistance, flow -volume loop, pressure-volume loop)
Monitoring of oxygen concentration:21%~100%
Security Alarm System
Oxygen Concentration alarm
upper limit setting range 21%~100%
low limit setting range 10%~80%
Airway pressure alarm
upper limit setting range 0.3~6.0 kPa
low limit setting range 0.2~5.0 kPa
Per-minute ventilation volume alarm
upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min
Sustained high-pressure alarm
it will give alarm when stress have consistently been higher than 2.5 kPa
Suffocation alarm:it will give sound and light alarm if there is no tidal volume input for 15 seconds
Power alarm
Power :AC 220V±10% 50Hz±1Hz. Optional UPS, Storage battery
Mechanical Arms
*Items within brackets ,Air compressor are optional.