LTSP17
EtCO2 Monitor (Capnograph)
Features
● Big screen LCD, clearly displays the EtCO2/CO2,Respiration rate
● EtCO2 Monitor (Capnograph)-LTM605B technology of infrared ray gas analysis:
○ Compensation system for temperature and pressure and Zero readjustment system
all these insure the accuracy and stability
○ Balance gas compensation makes minimum influence to other gases.
○ Lung pressure and temperature compensation (BTPS compensation) are selectable
○ Pump flux is selectable for adult and infant
● KPA , mmHg or % are selectable.
● Audio-optical alarm for exceeding limits and apnea
● 24 hours trend diagram keeps still available after power-off
● Mainstream mode more convenient, keeping work under the heavy humid environment, also suitable
for intubatton patient.
● Data output via RS-232 port
● Portable and lithium battery
● Digital SpO2 system optional accuracy can be insured even under the movement or
weak perfusion.
● Easy to connect and fix with anesthetic machine.
Technical data
Power
AC input:
100V-250V, 50Hz/60 Hz
AC power Consumption:
6.6W when battery is not being charged,
16W when battery is being charged
Operation
Working temperature:
10 -40°C (23 to 104°F)
Humidity:
30 -75% (non-condensing)
Atmospheric pressure:
86-106kPa
Battery Data
Type:
Lithium ion internal battery, rechargeable
Battery capacity:
12V,1800mAh, 8 hours (approximately)
Charging time:
4 hours from empty to full
Physical Data
Dimensions:
230mm(W)x160mm(H) x50mm(D)
Weight:
1150 grams
Capnography Measuring Data
EtCO2/CO2
Range:
0-99mmHg,or 0-13kPa,or 0-13%
Accuracy:
±2mmHg when 0-5%
±10%(reading) when > 5%
Update:
Once every breath
Trend length:
24hours
Respiration rate
Range:
3 -150 t/min
Accuracy:
the bigger value of ±10%(reading) or ±2 times/min
Trend length:
24 hours
PUMP Flow :
selectable from 50-250cc/min