Specifications
Test principle: Electrical resistence for counting and Cyanide free method for HGB.
Clot clear: High voltage burn and high pressure flush.
Sample volume: whole blood : 9.8μl, prediluted : 20μl
Aperture diameter: WBC: 100μ m, RBC、PLT: 70μ m
Chambers Two chambers.
Throughput: 60 samples/hour
Language English,Spanish,French,and Chinese.
Reference range type: newborn、children、male、female and general.
Display: 10.4 inch color touch screen, show all parameters and histograms
Input and output: USB、 keyboard、mouse、Parallel Port、RS232 serial port, can be connected to the network by external computer.
Working Environment: Temperature:10℃~35℃;Humidity:≤85% RH
Voltage: AC 110 V~240V,50Hz/60 Hz±1Hz ,Auto adapt according to the input voltage
Dimension : 370mm×435mm×472mm
Net weight: 30Kg
LT6400 vet differential blood cell counter Performance
Parameter | Linearity | Precision(CV%) |
WBC(109/L) | 0.3~99.9 | 2.5(7.0~15.0) |
RBC(102/L) | 0.20~8.0 | 2(3.5~6.0) |
HGB(g/L) | 10~250 | 1.5(110~180) |
MCV(FL) |
| 0.5(80.0~110.0) |
PLT(109/L) | 10~99.9 | 5(150~500) |
Carry over | WBC≤2% RBC≤1% HGB≤2% PLT≤2% |