LTSI16 Blood/Infusion/Feeding Warmer
  • LTSI16 Blood/Infusion/Feeding Warmer
  • LTSI16 Blood/Infusion/Feeding Warmer

  • No: LTSI16
  • Type: Syringe and infusion pump
  • Dimension:
  • Characteristics:

    Adapter Output: DC 15V, 2.0A


  • Details
  • Pictures
  • Service
  • Q&A
  • Compatible tube:Transfusion set, infusion set or feeding set with diameter 3.5-5mm

    Adapter Output: DC 15V, 2.0A

    Protection class:I

    Heating board Degree of protection:BF

    Main engine Protection against ingress of water:IPX2

    Control temperature:35°C - 42°C adjustable in 0.1°C steps

    Temperature accuracy:±0.5°C (heating plate), ±1°C (warming tube)

    Operational mode:Continuous operationDimensions :140x75x35mm Weight:210g

    service life:5 years

    Tube warmer Length: 800mm

    Permissible environmental conditions for transport and storage:Temperature:-20°C – 55°C

    Relative humidity:1090%

    Atmosphere range:70 – 106 Kpa

    Permissible environmental conditions for the ambient operating: 

    Temperature: +5  40°C.

    Values higher or lower than the ranges specified above may cause damage to the device or its accessories.

    Relative Humidity: ≤80%

    Atmosphere range: 70 – 106 Kpa

    Temperature

    Use the warming slot output reference table



    Infusion Rate(ml/h)

    25

    100

    180

    300

    600

    1000

    Slot Length

    Part

    All

    Part

    All

    Part

    All

    Part

    All

    Part

    All

    Part

    All

    Temperature Set

    35

    28.2

    29.0

    32.0

    33.0

    32.4

    33.2

    30.5

    32.0

    28.2

    29.4

    26.0

    27.0

     

    37

    29.2

    30.8

    33.4

    34.8

    33.4

    35.0

    31.6

    33.6

    28.8

    30.4

    26.5

    27.5

     

    42

    30.4

    33.2

    36.6

    39.0

    37.8

    39.5

    36.0

    37.5

    30.7

    34.4

    28.1

    30.0


    Note: Room temperature is 20 degrees, the temperature of the input liquid is 20 degrees, the output port of the warmer is at 125px from the outlet of the Warmer.

    Use warmer tube output reference table:


    Infusion Rate(ml/h)

    25

    100

    180

    300

    600

    1000

    Temperature Set

    35

    29.2

    33.8

    34.5

    34.8

    32.2

    30.0

     

    37

    31.2

    35.6

    36.5

    36.5

    33.8

    30.5

     

    42

    33.6

    39.8

    41.0

    41.0

    36.2

    32.5

    Note: Room temperature is 20 degrees, the temperature of the input liquid is 20 degrees, the output port of the warmer is at 125px from the outlet of the Warmer Tube.

    The heating effect is on the basis of the actual use (infusion environment temperature, infusion rate, soft and hard, thickness, and diameter of the infusion tube wall will affect the heating effect)