The technical parameters of LTSI03 infusion pump
1 infusion specifications: all kinds of domestic ordinary infusion and special silicone infusion
2 velocity selection: 1-1200 - ml/h (0.1 to 99.9 ml/h in 0.1 ml/h increase 100-1200 ml/h increase to 1 ml/h)
3 velocity accuracy: + / - 5%
4 dripping speed range: 1 / min - 333 drops/min
5 drops speed precision: monitoring range for 1-50 drops of precision
6: error compensation range 0-30% plus or minus 2% (interval)
7 transfusion volume Settings: 1-9999 - ml
8 have transfusion volume display range: 1-9999 - ml
9 KVO velocity: the speed of 1 ml/h infusion
10 bubbles detection: measurable bubble column length not less than 3 mm
11 drop sensor: infrared photoelectric detection beam
12 exhaust bubble: 800-999 ml/h, 1-9 ml could volume
13 BOLUS speed: 600 ml/h (500-800 ml/h is adjustable
Noise: 14 to 25 ml/h speed acuities were 36 db noise when running
15 alarm: jam, standby, bubbles, open the door, infusion, battery voltage, motor fault alarm
16 types: type I BF type
17 IP rating: IP x 4 splash water equipment
18 shows: large screen LCD
19 the dimensions: 13 X 10.2 X 20 cm
Power: 20 less than 20 va
21 weight: 1.3 KG
Two working conditions
1 power supply: AC/dc to AC, 100 v to 240 v, 50/60 hz
Built-in lithium battery, voltage DC12VC, fully charged for 4 hours, available for pump with 25 ml/h work 4 hours
2 environmental conditions: temperature 5 ℃ - + 40 ℃
Relative humidity 20% 90% RH or less
The atmospheric pressure: 860 hpa - 1060 hpa
Storage conditions: temperature 40 ℃ - + 55 ℃
Three functions: reserve a nurse call function, sound and light alarm function, standby time is adjustable, alarm sound attenuation, exhaust bubble function, ac/dc and dc power converter instructions, rapid transfusion control, flow rate drop infusion mode conversion, ac/dc switch alarm and indication, automatic tightening, infusion quantity is zero, KVO function, blocking release pills.